Fig. 4
- ID
- ZDB-IMAGE-261124-86
- Publication
- Hong et al., 2016 - A Novel Long-term, Multi-Channel and Non-invasive Electrophysiology Platform for Zebrafish
- All Figures
- Figures for Hong et al., 2016
Fig. 4
Long-term EEG monitoring capability of the iZAP system. Scn1Lab mutants of a chronic epilepsy model were restrained in the iZAP system between 3 to 8 dpf and continuously monitored with the surface electrodes.
(a,b) Representative electrograph and corresponding cross-correlation plot of 130 hours continuous monitoring. (c) 1-hour interval seizure score bar plot overlapped with circadian light cycle applied. (d,e) Zoom-ins of the electrograph and the time-derivative. At the maximal seizure activity at 28 hour, high-amplitude field potentials resulted in high-magnitude seizure scores while at the later stage like 8 dpf, high-amplitude field potentials were not related to seizure-like bursts. (f) Cumulative frequency plot for seizure-like bursts over a threshold of seizure score (seizure score >1). Three linear regions were found as indicated with α1, α2 and α3. Each corresponding R2 is indicated in the plot.